Dynamic
Balloon-Expandable Stent System
Indicated for the treatment of de novo or restenotic atherosclerotic lesions in iliac arteries.
Dynamic
Balloon-Expandable Stent System
Indicated for the treatment of de novo or restenotic atherosclerotic lesions in iliac arteries.
Excellent trackability
Stent designed for flexibility in iliac arteries
Improved stent surface biocompatibility
Excellent trackability
Stent designed for flexibility in iliac arteries
Improved stent surface biocompatibility
Technical data
Stent | |
---|---|
Stent material | Balloon-expandable |
Stent material | Stainless steel |
Strut thickness | 160 μm (ø 5.0 – 8.0 mm) |
180 μm (ø 9.0 – 10.0 mm) | |
Shortening | Negligible |
Stent coating | proBIO (Amorphous Silicon Carbide) |
Sizes | ø 5.0 – 10.0 mm; L: 15 – 25 – 38 – 56 mm |
Delivery System | |
---|---|
Catheter type | OTW |
Recommended guide catheter | 0.035″ |
Tip | Soft, short, tapered, colored |
Balloon makers | 2 swaged markers |
Shaft | 5 F, hydrophobic coating, dual-lumen |
Usable length | 80 and 130 cm (ø 5.0 – 8.0 mm) |
Markers | Two swaged markers |
Guide wire lumen | Hydrophobic coating |
Nominal pressure (NP) | 9 atm |
Rated burst pressure (RBP) | 15 atm (ø 5.0 – 8.0 mm) |
13 atm (ø 9.0 – 10.0 mm) |
Technical data
Stent | |
---|---|
Stent material | Balloon-expandable |
Stent material | Stainless steel |
Strut thickness | 160 μm (ø 5.0 – 8.0 mm) |
180 μm (ø 9.0 – 10.0 mm) | |
Shortening | Negligible |
Stent coating | proBIO (Amorphous Silicon Carbide) |
Sizes | ø 5.0 – 10.0 mm; L: 15 – 25 – 38 – 56 mm |
Delivery System | |
---|---|
Catheter type | OTW |
Recommended guide catheter | 0.035″ |
Tip | Soft, short, tapered, colored |
Balloon makers | 2 swaged markers |
Shaft | 5 F, hydrophobic coating, dual-lumen |
Usable length | 80 and 130 cm (ø 5.0 – 8.0 mm) |
Markers | Two swaged markers |
Guide wire lumen | Hydrophobic coating |
Nominal pressure (NP) | 9 atm |
Rated burst pressure (RBP) | 15 atm (ø 5.0 – 8.0 mm) |
13 atm (ø 9.0 – 10.0 mm) |